User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Amprobe RS 1000:Current Probe - Limited Cal DATE: 09-Nov-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 29 NUMBER OF LINES: 179 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A STANDARD: DTB DJK-50T ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F W 1.002 ASK+ X 1.003 HEAD * INITIAL CONDITIONS * 1.004 DISP The following equipment is need for this procedure: 1.004 DISP Toroid Coil DTB DJK DTB # 5-34 1.005 HEAD {* CURRENT TEST 15 AMP RANGE *} 1.006 JMP 2.001 1.007 EVAL 2.001 DISP [32] Connect the 5725A and the DJK01 as follows: 2.001 DISP [32] 5725A TO DJK COIL 2.001 DISP [32] CURRENT OUTPUT HI ......... + terminal 2.001 DISP [32] CURRENT OUTPUT LO ......... - terminal 2.002 STD DTB DJK-50T 2.003 DISP Set the UUT FUNCTION/RANGE switch to 15 AMP. 2.004 DISP Reading lock switch (next to function switch) should be 2.004 DISP in the left position. 2.005 DISP Clamp the UUT around the center of the coil. 2.006 DISP Adjust the UUT zero control (black adjustment on handel) 2.006 DISP for a UUT zero meter reading. 2.007 MMFC C3 2.008 MESS Adjust Stimulus for a UUT reading of 15 Amps. 2.009 5700 0.300A 60H N 2W 2.010 MEM* 50 2.011 MEMC 15 15.00A 5.00/ 60H #! Test Tol 0.75, Sys Tol 0.000235, TUR 3191.489 (>= 4.00). 3.001 MESS Adjust Stimulus for a UUT reading of 12 Amps. 3.002 5700 0.240A 60H N 2W 3.003 MEM* 50 3.004 MEMC 15 12.00A 5.00/ 60H #! Test Tol 0.75, Sys Tol 0.000196, TUR 3826.531 (>= 4.00). 4.001 MESS Adjust Stimulus for a UUT reading of 9 Amps. 4.002 5700 0.180A 60H N 2W 4.003 MEM* 50 4.004 MEMC 15 9.00A 5.00/ 60H #! Test Tol 0.75, Sys Tol 3.1e-005, TUR 24193.548 (>= 4.00). 5.001 MESS Adjust Stimulus for a UUT reading of 6 Amps. 5.002 5700 0.120A 60H N 2W 5.003 MEM* 50 5.004 MEMC 15 6.00A 5.00/ 60H #! Test Tol 0.75, Sys Tol 2.2e-005, TUR 34090.909 (>= 4.00). 6.001 MESS Adjust Stimulus for a UUT reading of 3 Amps. 6.002 5700 0.06A 60H N 2W 6.003 MEM* 50 6.004 MEMC 15 3.00A 5.00/ 60H #! Test Tol 0.75, Sys Tol 1.3e-005, TUR 57692.308 (>= 4.00). 7.001 HEAD {* CURRENT TEST 40 AMP RANGE *} 7.002 JMP 8.001 7.003 EVAL 8.001 DISP Set the UUT FUNCTION/RANGE switch to 40 AMP. 8.002 DISP Adjust the UUT zero control (black adjustment on handel) 8.002 DISP for a UUT zero meter reading. 8.003 MESS Adjust Stimulus for a UUT reading of 40 Amps. 8.004 5700 0.800A 60H N 2W 8.005 MEM* 50 8.006 MEMC 40 40.00A 5.00/ 60H #! Test Tol 2, Sys Tol 0.00056, TUR 3571.429 (>= 4.00). 9.001 MESS Adjust Stimulus for a UUT reading of 30 Amps. 9.002 5700 0.600A 60H N 2W 9.003 MEM* 50 9.004 MEMC 40 30.00A 5.00/ 60H #! Test Tol 2, Sys Tol 0.00043, TUR 4651.163 (>= 4.00). 10.001 MESS Adjust Stimulus for a UUT reading of 20 Amps. 10.002 5700 0.400A 60H N 2W 10.003 MEM* 50 10.004 MEMC 40 20.00A 5.00/ 60H #! Test Tol 2, Sys Tol 0.0003, TUR 6666.667 (>= 4.00). 11.001 MESS Adjust Stimulus for a UUT reading of 10 Amps. 11.002 5700 0.200A 60H N 2W 11.003 MEM* 50 11.004 MEMC 40 10.00A 5.00/ 60H #! Test Tol 2, Sys Tol 3.4e-005, TUR 58823.529 (>= 4.00). 12.001 HEAD {* CURRENT TEST 100 AMP RANGE *} 12.002 JMP 13.001 12.003 EVAL 13.001 DISP Set the UUT FUNCTION/RANGE switch to 100 AMP. 13.002 DISP Adjust the UUT zero control (black adjustment on handel) 13.002 DISP for a UUT zero meter reading. 13.003 MESS Adjust Stimulus for a UUT reading of 100 Amps. 13.004 5700 2A 60H N 2W 13.005 MEM* 50 13.006 MEMC 100 100.0A 5.00/ 60H #! Test Tol 5, Sys Tol 0.00134, TUR 3731.343 (>= 4.00). 14.001 MESS Adjust Stimulus for a UUT reading of 80 Amps. 14.002 5700 1.6A 60H N 2W 14.003 MEM* 50 14.004 MEMC 100 80.0A 5.00/ 60H #! Test Tol 5, Sys Tol 0.00108, TUR 4629.630 (>= 4.00). 15.001 MESS Adjust Stimulus for a UUT reading of 60 Amps. 15.002 5700 1.2A 60H N 2W 15.003 MEM* 50 15.004 MEMC 100 60.0A 5.00/ 60H #! Test Tol 5, Sys Tol 0.00082, TUR 6097.561 (>= 4.00). 16.001 MESS Adjust Stimulus for a UUT reading of 40 Amps. 16.002 5700 0.8A 60H N 2W 16.003 MEM* 50 16.004 MEMC 100 40.0A 5.00/ 60H #! Test Tol 5, Sys Tol 0.00056, TUR 8928.571 (>= 4.00). 17.001 MESS Adjust Stimulus for a UUT reading of 20 Amps. 17.002 5700 0.4A 60H N 2W 17.003 MEM* 50 17.004 MEMC 100 20.0A 5.00/ 60H #! Test Tol 5, Sys Tol 0.0003, TUR 16666.667 (>= 4.00). 18.001 HEAD {* CURRENT TEST 300 AMP RANGE *} 18.002 JMP 19.001 18.003 EVAL 19.001 DISP Set the UUT FUNCTION/RANGE switch to 300 AMP. 19.002 DISP Adjust the UUT zero control (black adjustment on handel) 19.002 DISP for a UUT zero meter reading. 19.003 MESS Adjust Stimulus for a UUT reading of 300 Amps. 19.004 5700 6A 60H B1 N 2W 19.005 MEM* 50 19.006 MEMC 300 300.0A 5.00/ 60H #! Test Tol 15, Sys Tol 0.00257, TUR 5836.576 (>= 4.00). 20.001 MESS Adjust Stimulus for a UUT reading of 250 Amps. 20.002 5700 5A 60H B1 N 2W 20.003 MEM* 50 20.004 MEMC 300 250.0A 5.00/ 60H #! Test Tol 15, Sys Tol 0.00217, TUR 6912.442 (>= 4.00). 21.001 MESS Adjust Stimulus for a UUT reading of 200 Amps. 21.002 5700 4A 60H B1 N 2W 21.003 MEM* 50 21.004 MEMC 300 200.0A 5.00/ 60H #! Test Tol 15, Sys Tol 0.00177, TUR 8474.576 (>= 4.00). 22.001 MESS Adjust Stimulus for a UUT reading of 150 Amps. 22.002 5700 3A 60H B1 N 2W 22.003 MEM* 50 22.004 MEMC 300 150.0A 5.00/ 60H #! Test Tol 15, Sys Tol 0.00137, TUR 10948.905 (>= 4.00). 23.001 MESS Adjust Stimulus for a UUT reading of 100 Amps. 23.002 5700 2A 60H B1 N 2W 23.003 MEM* 50 23.004 MEMC 300 100.0A 5.00/ 60H #! Test Tol 15, Sys Tol 0.00097, TUR 15463.918 (>= 4.00). 24.001 MESS Adjust Stimulus for a UUT reading of 50 Amps. 24.002 5700 1A 60H B1 N 2W 24.003 MEM* 50 24.004 MEMC 300 50.0A 5.00/ 60H #! Test Tol 15, Sys Tol 0.00057, TUR 26315.789 (>= 4.00). 25.001 HEAD {* CURRENT TEST 1k AMP RANGE *} 25.002 JMP 26.001 25.003 EVAL 26.001 DISP Set the UUT FUNCTION/RANGE switch to 1k AMP. 26.002 DISP Adjust the UUT zero control (black adjustment on handel) 26.002 DISP for a UUT zero meter reading. 26.003 MESS Adjust Stimulus for a UUT reading of 500 Amps. 26.004 5700 10A 60H B1 N 2W 26.005 MEM* 50 26.006 MEMC 1000 500.0A 5.00/ 60H #! Test Tol 50, Sys Tol 0.00417, TUR 11990.408 (>= 4.00). 27.001 MESS Adjust Stimulus for a UUT reading of 400 Amps. 27.002 5700 8A 60H B1 N 2W 27.003 MEM* 50 27.004 MEMC 1000 400.0A 5.00/ 60H #! Test Tol 50, Sys Tol 0.00337, TUR 14836.795 (>= 4.00). 28.001 MESS Adjust Stimulus for a UUT reading of 200 Amps. 28.002 5700 4A 60H B1 N 2W 28.003 MEM* 50 28.004 MEMC 1000 200.0A 5.00/ 60H #! Test Tol 50, Sys Tol 0.00177, TUR 28248.588 (>= 4.00). 29.001 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.